Chromatography: principles, paper, thin layer, column, gas-liquid

Analytical Techniques – Chromatography (Cambridge IGCSE/A‑Level Chemistry 9701, Topic 22)

0. Context within the Syllabus

Chromatography is one of the major analytical families studied alongside:

  • Spectroscopic methods – UV‑Vis, IR, NMR, atomic absorption.
  • Electro‑chemical techniques – potentiometry, voltammetry.
  • Classical quantitative methods – gravimetry, titrimetry.

Examination objectives (AO1–AO3) require candidates to:

  • Select the most appropriate technique for a given sample.
  • Explain the underlying principles (partition, selectivity, resolution, etc.).
  • Interpret qualitative data (Rf, retention time) and perform quantitative calculations (calibration curves, % recovery).
  • Design, carry out and critically evaluate a chromatographic experiment.

Decision‑making flow‑chart (sample → technique)

Sample descriptionPreferred techniqueRationale
Non‑volatile, solid mixture; only a few mg available Paper chromatography or TLC Simple, inexpensive, works with small amounts; gives Rf values for identification.
Non‑volatile, larger quantity, need to isolate product Column chromatography (gravity or flash) Scalable, can collect purified fractions.
Volatile, thermally stable compounds (e.g., hydrocarbons, alcohols) Gas‑liquid chromatography (GC) Fast, highly sensitive, provides quantitative data via peak areas.
Mixture of polar & non‑polar compounds, need high resolution High‑performance liquid chromatography (HPLC) – not covered in depth for IGCSE but relevant for A‑Level. Reverse‑phase column gives excellent separation of a wide polarity range.

1. Fundamental Principles of Chromatography

  • Stationary phase: solid (silica, alumina, cellulose) or a liquid film on a solid.
  • Mobile phase: liquid (paper, TLC, column) or gas (GC).
  • Partition coefficient (K) – equilibrium distribution of a solute between stationary and mobile phases: $$K=\frac{C_{\text{stationary}}}{C_{\text{mobile}}}$$
  • Selectivity (α) – ability of the system to distinguish two components: $$\alpha=\frac{K_2}{K_1}\qquad(K_2>K_1)$$
  • Retention factor (Rf) – planar chromatography**: $$R_f=\frac{\text{distance travelled by component}}{\text{distance travelled by solvent front}}$$
  • Retention time (tR) – column chromatography**: time taken for a component to elute from the column.
  • Resolution (Rs)** – separation between two adjacent peaks: $$R_s=\frac{2\,(t_{R2}-t_{R1})}{w_1+w_2}$$ where $w$ = peak width at base.

2. Paper Chromatography

2.1 Principle & Set‑up

  • Stationary phase: cellulose paper (highly polar).
  • Mobile phase: organic solvent (commonly butanol : acetone = 4 : 1).
  • Separation is based on differential adsorption of components onto the polar paper versus dissolution in the mobile solvent.

2.2 Procedure

  1. Draw a faint pencil line ≈2 mm from the bottom edge of a strip of chromatography paper.
  2. Apply a ≤2 µL spot of the sample solution onto the line; allow it to dry.
  3. Place the strip in a sealed developing chamber containing a shallow pool of solvent (solvent level < spot).
  4. Seal the chamber; the solvent rises by capillary action. When the front reaches ~80 % of the strip length, remove the paper.
  5. Immediately mark the solvent front with a pencil.
  6. Visualise the spots (UV lamp, iodine vapour, or a specific staining reagent).
  7. Measure distances (to 0.1 cm) and calculate Rf values.

2.3 Interpretation (Qualitative Identification)

Compare the measured Rf values with a reference table obtained under identical conditions.

CompoundSolvent system (butanol : acetone = 4 : 1)Typical Rf
Glucose0.45
Fructose0.62
Sucrose0.78

Worked example (exam style):

A paper chromatogram of an unknown sugar gives Rf = 0.46. Which sugar is most likely present?

Solution: The value matches glucose (Rf ≈ 0.45). Hence the unknown is identified as glucose.

3. Thin‑Layer Chromatography (TLC)

3.1 Principle & Materials

  • Stationary phase: a thin layer (≈0.2–0.3 mm) of silica gel or alumina coated on a glass, aluminium or plastic plate.
  • Mobile phase: an organic solvent or mixture (e.g., ethyl acetate : hexane = 1 : 3).
  • Separation occurs by differential adsorption of the analytes onto the polar adsorbent versus their solubility in the mobile solvent.

3.2 Procedure

  1. Draw a faint line 1 cm from the bottom edge of the plate.
  2. Spot ≤2 µL of the sample onto the line using a capillary tube; allow to dry.
  3. Place the plate in a sealed developing chamber containing a shallow pool of solvent (solvent level < spot).
  4. Develop until the solvent front is ~8 cm from the origin.
  5. Remove, dry, and visualise under UV (254 nm) or by iodine vapour, ninhydrin, etc.
  6. Measure distances and calculate Rf values.

3.3 Advantages over Paper Chromatography

  • Higher resolution – thinner stationary layer reduces band broadening.
  • Rapid development (typically 5–10 min).
  • Wider choice of solvents, including non‑polar mixtures.
  • Quantitative analysis possible via densitometry (spot‑intensity measurement).

3.4 Example

Separate a mixture of benzoic acid, acetophenone and phenol using silica TLC with ethyl acetate : hexane = 1 : 3. Measured Rf values: 0.21 (benzoic acid), 0.45 (acetophenone), 0.78 (phenol). The values are within the optimal 0.2–0.8 range, confirming a suitable solvent system.

4. Column Chromatography (Preparative)

4.1 Types of Columns

FeatureNormal‑phase columnReverse‑phase column
Stationary phasePolar silica gel or aluminaNon‑polar C18‑bonded silica
Mobile phaseNon‑polar solvents (hexane, petroleum ether)Polar solvents (water, methanol, acetonitrile)
Typical useSeparation of non‑polar to moderately polar compoundsSeparation of polar to moderately non‑polar compounds; basis of HPLC

4.2 Column Packing (wet‑packing method)

  1. Choose column dimensions (e.g., 2 cm × 30 cm). Approx. 30 g silica per 10 cm column length.
  2. Slurry the silica in the chosen solvent (e.g., hexane) to form a uniform suspension.
  3. Pour the slurry into the column, allowing it to settle without forming air bubbles.
  4. Tap gently to settle the bed; add more slurry until the column is filled to the desired height.
  5. Condition the packed column by passing 5–10 column volumes of the mobile phase.

4.3 Running the Column

  1. Apply the sample on top of the stationary phase (dry‑load for solid samples or dissolve in a minimal amount of solvent).
  2. Elute with the mobile phase at a steady flow (gravity or low‑pressure pump).
  3. Collect fractions of a convenient volume (e.g., 10 mL).
  4. Analyse each fraction by TLC or GC to locate the target component.
  5. Combine the appropriate fractions and remove solvent (rotary evaporator).

4.4 Quantitative Example – % Recovery

A 0.500 g mixture contains 0.120 g of component A. After column chromatography 0.108 g of A is isolated.

$$\%\,\text{recovery}= \frac{0.108\;\text{g}}{0.120\;\text{g}}\times100 = 90\%$$

5. Gas‑Liquid Chromatography (GC)

5.1 Principle & Main Components

  • Stationary phase: a high‑boiling liquid (e.g., 5 % phenyl‑methylpolysiloxane) coated on the inner wall of a fused‑silica capillary (10–30 m long, 0.1–0.53 mm i.d.).
  • Mobile phase: inert carrier gas (helium or nitrogen) at a constant flow (≈ 1 mL min⁻¹).
  • Detector: flame ionisation detector (FID), thermal conductivity detector (TCD) or mass spectrometer (GC‑MS).
  • Separation arises from repeated partition of vapourised analytes between the gas phase and the liquid film.

5.2 Typical Run (step‑by‑step)

  1. Inject ~1 µL of liquid sample into a heated injector (220–250 °C for most organics).
  2. The sample instantly vapourises and is carried onto the column by the carrier gas.
  3. Components interact with the stationary liquid film; stronger interaction → longer retention time.
  4. Each component elutes at a characteristic tR and produces a peak on the chromatogram.
  5. Peak area is proportional to the amount of component present.

5.3 Key Parameters

  • Column dimensions – longer columns give higher efficiency; smaller i.d. improves resolution but increases back‑pressure.
  • Temperature programming – isothermal for narrow‑boiling‑range mixtures; ramped (e.g., 40 °C → 250 °C at 10 °C min⁻¹) for complex samples.
  • Resolution (Rs)** – see equation in Section 1.

5.4 Quantitative Example – Calibration Curve

  1. Prepare standards of a volatile compound (e.g., n‑butanol) at 0.1, 0.5, 1.0 and 2.0 mg mL⁻¹.
  2. Inject each under identical GC conditions; record peak areas (A).
  3. Plot A (y‑axis) against concentration (C, x‑axis). A straight line is obtained: $$A = 1.23\times10^{5}\,C + 2.1\times10^{3}$$
  4. Inject an unknown; measured area = 6.18 × 10⁴. $$C_{\text{unk}} = \frac{6.18\times10^{4} - 2.1\times10^{3}}{1.23\times10^{5}} = 0.49\;\text{mg mL}^{-1}$$

6. Quantitative Chromatography – General Approach

6.1 Calibration Curves (any technique)

  1. Prepare a series of standards covering the expected concentration range.
  2. Analyse each standard under identical conditions; record the analytical response (peak area, intensity, etc.).
  3. Plot response (y) vs. concentration (x) and obtain a linear equation $y = mx + c$.
  4. Measure the response of the unknown and calculate its concentration using the equation.

6.2 % Recovery (Preparative work)

$$\%\,\text{recovery}= \frac{m_{\text{recovered}}}{m_{\text{theoretical}}}\times100$$

Useful for evaluating the efficiency of a column or extraction procedure.

7. Practical Skills, Experimental Design & Safety

7.1 Designing a Chromatographic Experiment (AO3)

  1. Define the aim – e.g., “Separate and identify the components of a mixed food‑colour sample”.
  2. Choose the technique – start with TLC for rapid screening; if components are volatile, move to GC.
  3. Select a solvent system – test at least two ratios; target Rf (or tR) values between 0.2 and 0.8 (or spacing > 1 min in GC).
  4. Plan controls – run known standards alongside the unknown.
  5. Determine data handling – record distances to 0.1 cm, calculate Rf to two decimal places; for GC, record tR to 0.01 min and integrate peaks with software.
  6. Evaluate results – discuss possible sources of error (solvent purity, temperature fluctuations, column packing) and suggest improvements.

7.2 Safety Considerations

HazardPrecaution
Organic solvents (hexane, acetone, butanol) Use a fume hood, wear gloves and safety goggles, keep away from ignition sources.
High‑temperature injector & column (GC) Allow equipment to cool before servicing; use heat‑resistant gloves when handling liners.
Silica/alumina dust (column packing) Wear a dust mask, avoid inhalation, and clean spills immediately.
Pressurised systems (flash column, HPLC) Never seal a column completely; use pressure‑relief valves and check fittings before use.

8. Summary Comparison of Chromatographic Techniques

Technique Stationary Phase Mobile Phase Typical Sample Type Key Advantages Key Limitations
Paper Chromatography Cellulose (polar) Organic liquid Small, soluble mixtures (pigments, sugars) Very simple, inexpensive, visual Low resolution; limited to soluble, non‑volatile samples
Thin‑Layer Chromatography (TLC) Silica gel / alumina (adsorbent) Organic solvent or mixture Organic compounds, pharmaceuticals, pigments Fast, high resolution, quantitative (densitometry) Limited sample load; requires careful spotting
Column Chromatography Silica, alumina or bonded phases (normal or reverse) Liquid solvent (gravity or pressure) Mixtures needing preparative separation Scalable, can be automated (flash, HPLC) Longer run times, higher solvent consumption
Gas‑Liquid Chromatography (GC) Liquid film on fused‑silica capillary Inert gas (He, N₂) Volatile, thermally stable compounds High sensitivity, rapid, excellent quantification Unsuitable for non‑volatile or thermally labile substances

9. Practical Tips for A‑Level Examinations

  • Always mark the solvent front immediately after removal and calculate Rf to two decimal places.
  • Select a solvent system that gives Rf values between 0.2 and 0.8 for all components of interest.
  • When packing a column, ensure the bed is uniform and free of air bubbles – tap gently and allow the slurry to settle.
  • For GC, set the injector temperature just above the boiling point of the sample; avoid excessive heating that can cause decomposition.
  • In data interpretation, compare Rf or tR values with those of standards run under identical conditions.
  • When constructing a calibration curve, use at least four concentration points and check that the correlation coefficient (R²) is ≥ 0.99.
  • Discuss sources of error (e.g., solvent evaporation, incomplete elution, detector drift) and suggest realistic improvements.

Create an account or Login to take a Quiz

28 views
0 improvement suggestions

Log in to suggest improvements to this note.