understand the use of the time-base and y-gain of a cathode-ray oscilloscope (CRO) to determine frequency and amplitude

Progressive Waves – Using a Cathode‑Ray Oscilloscope (CRO)

1. Aim

To use the horizontal time‑base and vertical y‑gain controls of a CRO to obtain the frequency and amplitude of a travelling (progressive) wave, and to relate these measurements to the quantities required by the Cambridge 9702 – Waves syllabus (7.1‑7.3).

2. Key Wave Quantities (Syllabus 7.1)

QuantityDefinition (textbook)Symbol in notes
DisplacementInstantaneous distance of a particle from its equilibrium position.y (or A for maximum displacement)
Amplitude (A)Maximum displacement from equilibrium.A
Phase (ϕ)Fraction of a complete cycle at a given instant (rad or °).ϕ
Period (T)Time for one complete cycle.T = 1/f
Frequency (f)Number of cycles per second.f = 1/T
Wavelength (λ)Distance between two successive points in phase (e.g. two peaks).λ
Wave speed (v)Speed at which a given phase propagates.v = f λ
Intensity (I)Power transmitted per unit area (I = P/A). For a string I = ½ ρ v A².I

3. CRO Controls Relevant to Wave Measurements

ControlSymbolPhysical meaningTypical units
Time‑Base (Horizontal)SSweep speed – horizontal distance travelled by the electron beam per unit time.seconds per division (s/div)
Y‑Gain (Vertical)VdivVoltage sensitivity – vertical displacement per unit voltage.volts per division (V/div)

4. Determining Frequency from the Time‑Base

  1. Display a sinusoidal voltage on the screen.
  2. Count the number of horizontal divisions that contain one complete cycle – call this Lh (div).
  3. Calculate the period:

    \$T = L_{h}\,S\$

  4. Frequency follows directly:

    \$f = \frac{1}{T}= \frac{1}{L_{h}\,S}\$

5. Determining Amplitude from the Y‑Gain

  1. Measure the vertical peak‑to‑peak height in divisions – Lv (div).
  2. Convert to peak‑to‑peak voltage:

    \$V{pp}=L{v}\,V_{div}\$

  3. For a sinusoid the amplitude (peak value) is half the peak‑to‑peak voltage:

    \$A = \frac{V{pp}}{2}= \frac{L{v}\,V_{div}}{2}\$

6. From Frequency and Wavelength to Wave Speed (v = f λ)

The CRO gives the frequency of the electrical signal that is produced by the mechanical wave source. The wavelength must be obtained from a physical measurement on the medium.

6.1 Measuring the Wavelength

  • Node‑spacing method (standing wave on a string): distance between two adjacent nodes = λ/2 →

    \$\lambda = 2\,(\text{node spacing})\$

  • Mode‑number method: if the string length is L and the standing‑wave mode number is n (number of half‑wavelengths), then

    \$\lambda = \frac{2L}{n}\$

6.2 Derivation of v = f λ using CRO data

From the CRO we have measured f. The wavelength is obtained as described above. Substituting the two measured quantities gives the wave speed:

\$v = f\,\lambda = \frac{1}{L_{h}S}\;\times\;\lambda\$

This explicit link – “frequency from the time‑base, wavelength from the medium → wave speed” – satisfies the syllabus requirement that students *show* how the CRO data lead to the wave‑speed equation.

7. Intensity and Its Dependence on Amplitude

For a transverse wave on a string the intensity (average power per unit cross‑sectional area) is defined by the syllabus as

\$I = \frac{P}{A_{\text{cross}}}\$

where P is the average power transmitted and Across the cross‑sectional area of the string. Using the standard result for a sinusoidal travelling wave on a stretched string, the average power is

\$P = \tfrac{1}{2}\,\rho\,v\,A^{2}\$

Dividing by the (constant) cross‑sectional area gives the intensity formula used in the notes:

\$I = \frac{1}{2}\,\rho\,v\,A^{2}\qquad\text{(∝ A²)}\$

Thus, any change in the measured vertical division count Lv (and therefore in A) leads to a quadratic change in intensity – a point that often appears in AO2 exam questions.

8. Uncertainty Propagation (AO2)

When reporting results, include the uncertainties arising from reading divisions on the CRO and from the physical measurement of wavelength.

  • Horizontal division count: ΔLh ≈ ±0.1 div
  • Vertical division count: ΔLv ≈ ±0.1 div
  • Time‑base setting: ΔS (usually given by the instrument, e.g. ±1 % of the selected value)
  • Y‑gain setting: ΔVdiv (similarly ±1 %)
  • Wavelength (node spacing or mode‑number method): Δλ from ruler accuracy (±0.5 mm) or from the uncertainty in n.

Using standard propagation rules:

\[

\frac{\Delta f}{f}= \sqrt{\left(\frac{\Delta Lh}{Lh}\right)^2+\left(\frac{\Delta S}{S}\right)^2},

\qquad

\frac{\Delta A}{A}= \sqrt{\left(\frac{\Delta Lv}{Lv}\right)^2+\left(\frac{\Delta V{div}}{V{div}}\right)^2},

\]

\[

\frac{\Delta v}{v}= \sqrt{\left(\frac{\Delta f}{f}\right)^2+\left(\frac{\Delta \lambda}{\lambda}\right)^2},

\qquad

\frac{\Delta I}{I}= \sqrt{\left(\frac{\Delta \rho}{\rho}\right)^2+\left(\frac{\Delta v}{v}\right)^2+4\left(\frac{\Delta A}{A}\right)^2}.

\]

9. Transverse vs. Longitudinal Waves (Syllabus 7.2)

AspectTransverse waveLongitudinal wave
Particle displacementPerpendicular to direction of propagationParallel to direction of propagation
Typical mediumStretched string, surface water, EM wave (electric field)Sound in air or gas, compression wave in a spring
Common transducer for CROPickup coil, photodiode, or tension‑meter that converts vertical motion to voltageMicrophone or pressure‑sensitive cell that converts pressure variations to voltage
Signal shown on CROVoltage proportional to transverse displacement (sinusoidal if the mechanical wave is sinusoidal)Voltage proportional to pressure variation (also sinusoidal for a pure tone)

10. Doppler Effect (Syllabus 7.3 – optional extension)

If the source moves with speed vs relative to the medium, the observed frequency on the CRO changes to

\$f{o}=f{s}\,\frac{v}{v\pm v_{s}}\$

  • Use “+” when the source moves away from the observer, “–” when it moves towards the observer.
  • The CRO directly displays the shifted frequency fo; rearranging the formula yields the original source frequency fs.

11. Worked Example (All Quantities)

Given

  • Time‑base S = 0.5 ms/div
  • Y‑gain Vdiv = 2 V/div
  • Horizontal length of one cycle Lh = 4 div
  • Vertical peak‑to‑peak height Lv = 3 div
  • String length L = 0.80 m; node spacing = 0.10 m → λ = 0.20 m
  • Linear mass density ρ = 0.02 kg m⁻¹
  • Transducer conversion: 1 V ↔ 0.5 mm displacement

  1. Period:  T = 4 div × 0.5 ms/div = 2 ms
  2. Frequency:  f = 1/T = 500 Hz
  3. Peak‑to‑peak voltage:  Vpp = 3 div × 2 V/div = 6 V
  4. Amplitude (electrical):  Ae = 6 V/2 = 3 V

    Mechanical amplitude:  A = 3 V × 0.5 mm V⁻¹ = 1.5 mm = 1.5 × 10⁻³ m

  5. Wavelength:  λ = 2 × 0.10 m = 0.20 m
  6. Wave speed:  v = f λ = 500 Hz × 0.20 m = 100 m s⁻¹
  7. Intensity:  I = ½ ρ v A² = ½ (0.02)(100)(1.5 × 10⁻³)² ≈ 2.3 × 10⁻⁶ W m⁻¹

12. Practical Tips (AO1 & AO2)

  • Calibrate the CRO with the built‑in calibration signal before taking data.
  • Adjust the time‑base so that one full cycle occupies 2–4 horizontal divisions – this minimises reading error.
  • Set the y‑gain so the waveform uses about 2–4 vertical divisions peak‑to‑peak without clipping.
  • Use the CRO’s cursors or the “measure” function for division counts; typical reading uncertainty ±0.1 div.
  • When measuring wavelength on a string, keep the tension constant and use a ruler with millimetre accuracy.
  • Record the uncertainties of S, Vdiv, Lh, Lv and λ; propagate them as shown in Section 8.

13. Summary Table

QuantityMeasured on CROFormulaUnits
Period (T)Horizontal divisions LhT = Lh Ss
Frequency (f)Period Tf = 1/THz
Peak‑to‑peak voltage (Vpp)Vertical divisions LvVpp = Lv VdivV
Amplitude (A)VppA = Vpp/2V (electrical) or m (mechanical after conversion)
Wavelength (λ)Physical measurement on the mediumλ = 2 × (node spacing)  or  λ = 2L/nm
Wave speed (v)f and λv = f λm s⁻¹
Intensity (I)Amplitude A (from CRO)I = ½ ρ v A²W m⁻¹

Typical CRO display: one full sinusoidal cycle with labelled horizontal (time) and vertical (voltage) divisions; inset shows a string with a pickup coil feeding the CRO.